Hi, I'm empty!
Please fill me with a spirometer, PFT filters, or a calibration syringe.
Astra BT is a wireless spirometer functions with two AAA batteries and offers mobility for occupational health professionals as well as for primary care clinics and hospitals. The spirometer comes with the AstraPro software.
More...Astra BT is a wireless spirometer functions with two AAA batteries and offers mobility for occupational health professionals as well as for primary care clinics and hospitals. The spirometer comes with the AstraPro software.
Features:
AccuFlo™ Digital Turbine
The AccuFlo™ bidirectional turbine with optoelectronic rotary sensor can be detached for cleaning. Sapphire bearings ensure optimal performance for the life of the turbine. AccuFlo has a unique flow trigger algorithm that allows tidal breathing prior to the patient expiratory effort making the spirometry test easier for the patient and insuring quality test results.
AstraPro™ Software
Choose either pre- and post-bronchodilator and bronchial challenge testing or use the incentive program with your pediatric patients. The Astra BT communicates with AstraPro wirelessly via Bluetooth.
Astra BT Spirometer Specifications | |
Forced Parameters Measured | |
Forced vital capacity: | FVC, FEV1, FEV1/FVC, PEF, FEF50%, FEV6, FEV1/FEC, PEF, FEF25-75%, FEV6, FEV1/ FEV0.5, PEFT, Vext, FIVC, FIF50%, FEF50/FIF50, QC Grades, FEV3, FEV0.5, FEV0.5/FVC, FEV3/FVC, FEV1/VC, FEV1/FEV6, FEV1/PEF, FEV1/FIV1, PEF/PIF, FEV25%, FEF25%, FEF75%, FEF75-85%, FET25-75%, FET100, FIV1, FIV1/FVC, PIF, MTT, MVV Ind, Lung age |
MVV: | MVV, Br./min |
Bronchoconstriction: | FVC, FEV1, PEF, FEF25-75%, PDX |
Relaxed Parameters Measured | |
VC: | VC, TV, ERV, IRV, IC, Ti, Te, Tt, Ti/Tt |
Maneuvers per patient: | 8 Total; Best test and last test highlighted |
Interpretation: | Kory, Miller, ATS, NLHEP (Ferguson) |
Predicted sets: | GLI/QUANJER, SEPAR 2013, ERS, KNUDSON, CRAPO, ZAPLETAL, MORRIS, AUSTRIA, GUTIERREZ, CASTRO - PEREIRA 2002, POLGAR-WENG, HANKINSON-NHANES III, PEREZ-PADILLA, A.J. CRUZ GOLSHAN, GARCIA RIO, CANDELA, PLATINO, THAI 2000, CASTRO PEREIRA 2007 |
Ethnic correction: | % Actual, Caucasian, Hispanic, African-American |
Language: | English, Spanish, French |
Measurement Scale (BTPS) | |
Flow range: | 0–16 L/sec |
Volume: | 0–10 L |
Measurement method: | Infrared Interruption via Digital Turbine |
Dynamic flow resistance: | < 1.5 cm H20/L/sec @ 14 L/sec |
Measurement Precision (BTPS) | |
At highest volume: | ± 3% or 50 ml |
At highest flow: | ± 5% or 150 ml |
Time related precision: | 0.5% |
Volume resolution: | < 6 mL |
Sampling frequency: | 100 Hz |
Power, Ports, and Dimensions | |
Batteries | 2 Alkaline AAA batteries, or 2 rechargeable AAA batteries |
Connectivity: | Bluetooth 2.0, USB 2.0, USB 3.0 |
Weight: | 6 oz (173 g) |
Dimensions: | 7.1 x 3.5 x 2.0 in (18 x 9 x 5 cm) |